• DocumentCode
    1346242
  • Title

    Low-Power Wireless Micromanometer System for Acute and Chronic Bladder-Pressure Monitoring

  • Author

    Majerus, Steve J A ; Fletter, Paul C. ; Damaser, Margot S. ; Garverick, Steven L.

  • Author_Institution
    Dept. of Electr. Eng. & Comput. Sci., Case Western Reserve Univ., Cleveland, OH, USA
  • Volume
    58
  • Issue
    3
  • fYear
    2011
  • fDate
    3/1/2011 12:00:00 AM
  • Firstpage
    763
  • Lastpage
    767
  • Abstract
    This letter describes the design, fabrication, and testing of a wireless bladder-pressure-sensing system for chronic, point-of-care applications, such as urodynamics or closed-loop neuromodulation. The system consists of a miniature implantable device and an external RF receiver and wireless battery charger. The implant is small enough to be cystoscopically implanted within the bladder wall, where it is securely held and shielded from the urine stream. The implant consists of a custom application-specific integrated circuit (ASIC), a pressure transducer, a rechargeable battery, and wireless telemetry and recharging antennas. The ASIC includes instrumentation, wireless transmission, and power-management circuitry, and on an average draws less than 9 μA from the 3.6-V battery. The battery charge can be wirelessly replenished with daily 6-h recharge periods that can occur during the periods of sleep. Acute in vivo evaluation of the pressure-sensing system in canine models has demonstrated that the system can accurately capture lumen pressure from a submucosal implant location.
  • Keywords
    application specific integrated circuits; biomedical electronics; biomedical measurement; low-power electronics; patient monitoring; pressure measurement; prosthetics; wireless channels; application-specific integrated circuit; bladder wall; chronic bladder-pressure monitoring; closed-loop neuromodulation; external RF receiver; low-power wireless micromanometer system acute bladder-pressure monitoring; lumen pressure; miniature implantable device; point-of-care application; power management circuitry; pressure transducer; rechargeable battery; recharging antennas; submucosal implant location; urine stream; urodynamics; wireless battery charger; wireless telemetry; wireless transmission; Application specific integrated circuits; Batteries; Bladder; Implants; Telemetry; Wireless communication; Wireless sensor networks; Bladder pressure; implantable biomedical device; low-power implantable device; micromanometer; wireless implant; wireless recharging; Animals; Dogs; Electronics, Medical; Humans; Manometry; Models, Biological; Monitoring, Physiologic; Pressure; Prostheses and Implants; Swine; Telemetry; Urinary Bladder;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.2010.2085002
  • Filename
    5597939