Title :
Development of 7.75 Ratio Voltage Divider Toward a Precise Measurement of Decade Resistance Based on the AC Quantized Hall Resistance
Author :
Domae, Atsushi ; Sakamoto, Norihiko ; Kiryu, Syogo ; Kaneko, Nobu-hisa
Author_Institution :
Nat. Metrol. Inst. of Japan, Nat. Inst. of Adv. Ind. Sci. & Technol., Tsukuba, Japan
Abstract :
At the National Metrology Institute of Japan, we are developing a transformer bridge circuit that can precisely compare the i = 2 plateau resistance in an ac quantized Hall resistance with a 100-kQ resistance. In this paper, we propose a practical transformer voltage ratio of 7.75:1, which is the most important parameter for developing a transformer bridge circuit. Further, we describe the design of an inductive voltage divider (IVD) to realize the 7.75:1 ratio. We intend to use the proposed IVD as a reference standard for calibrating the voltage ratio of the transformer. The reference IVD was fabricated with the two-stage technique and calibrated by Thompson´s method. The deviation from the nominal ratio of 7.75 was estimated to be -5.92×10-6 for in-phase and -22.75×10-6 for quadrature. The standard uncertainty of this calibration was estimated to be 0.63 × 10-6 for in-phase and 6.28 × 10-6 for quadrature.
Keywords :
Hall effect; bridge circuits; calibration; electric resistance measurement; measurement standards; voltage dividers; ac quantized Hall resistance; calibration; decade resistance measurement; inductive voltage divider; transformer bridge circuit; transformer voltage ratio; Bridge circuits; Calibration; Resistance; Resistors; Standards; Voltage measurement; Windings; AC quantized Hall resistance (QHR); ac resistance; bridge circuit; inductive voltage divider (IVD); voltage ratio; voltage ratio.;
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
DOI :
10.1109/TIM.2014.2383131