• DocumentCode
    1366872
  • Title

    The new charged-particle veto detector for the photon spectrometer TAPS

  • Author

    Janssen, S. ; Döring, W. ; Metag, V. ; Novotny, R.

  • Author_Institution
    II. Phys. Inst., Giessen Univ., Germany
  • Volume
    47
  • Issue
    3
  • fYear
    2000
  • fDate
    6/1/2000 12:00:00 AM
  • Firstpage
    798
  • Lastpage
    801
  • Abstract
    A new charged-particle veto detector for the photon spectrometer TAPS has been designed, tested and manufactured. The hexagonal plastic scintillators of 5 mm thickness are read-out via green emitting wavelength-shifting fibers and multi-anode photomultipliers. The concept, the performance of prototypes tested with high energy electrons and cosmic muons and the final design are presented. The limitations of fast and compact multi-anode metal case photomultipliers are discussed in detail
  • Keywords
    electron detection; gamma-ray detection; gamma-ray spectrometers; muon detection; photomultipliers; solid scintillation detectors; TAPS; charged-particle veto detector; compact multi-anode metal case photomultipliers; cosmic muons; green emitting wavelength-shifting fibers; hexagonal plastic scintillators; high energy electrons; multi-anode photomultipliers; photon spectrometer; Assembly; Manufacturing; Mesons; Optical fiber testing; Optical fibers; Photomultipliers; Plastics; Radiation detectors; Solid scintillation detectors; Spectroscopy;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/23.856519
  • Filename
    856519