DocumentCode
1367513
Title
Vibrostability of high pressure xenon gamma-ray detectors
Author
Dmitrenko, V.V. ; Chernysheva, I.V. ; Gratchev, V.M. ; Kondakova, O.N. ; Krivova, K.V. ; Ulin, S.E. ; Uteshev, Z.M. ; Vlasik, K.F.
Author_Institution
Eng. & Phys. Inst., Acad. of Sci., Moscow, Russia
Volume
47
Issue
3
fYear
2000
fDate
6/1/2000 12:00:00 AM
Firstpage
939
Lastpage
943
Abstract
For about 20 years high-pressure xenon (HPXe) detectors have been developed for the measurement of gamma-radiation spectra in the energy range 0.1-5 MeV with high-energy resolution. These detectors have numerous advantages in comparison with germanium and scintillation detectors, but they are relatively sensitive to the influence of acoustic and mechanical vibrations. Decreasing the microphonic sensitivity is necessary for improving the characteristics of these detectors. In this paper the results of acoustic and vibration testing performed on parallel-plate and cylindrical HPXe detectors are presented. Design modifications of high-pressure xenon detectors for improved vibration stability are discussed
Keywords
gamma-ray detection; ionisation chambers; vibrations; xenon; 0.1 to 5 MeV; Xe; gamma-radiation spectra; high pressure xenon gamma-ray detectors; mechanical vibrations; microphonic sensitivity; vibration stability; vibrostability; Acoustic measurements; Acoustic signal detection; Energy measurement; Energy resolution; Gamma ray detection; Gamma ray detectors; Germanium; Solid scintillation detectors; Vibrations; Xenon;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/23.856722
Filename
856722
Link To Document