DocumentCode
137723
Title
Preliminary evaluation of a new microsurgical robotic system for head and neck surgery
Author
Olds, Kevin C. ; Chalasani, Preetham ; Pacheco-Lopez, Paulette ; Iordachita, Iulian ; Akst, Lee M. ; Taylor, Russell H.
Author_Institution
Johns Hopkins Univ., Baltimore, MD, USA
fYear
2014
fDate
14-18 Sept. 2014
Firstpage
1276
Lastpage
1281
Abstract
This paper presents an implementation and evaluation of the Robotic ENT Microsurgery System (REMS). The implementation is discussed in reference to analysis from previous work, and evaluated using a simulated surgical task designed to resemble microlaryngeal phonosurgery. Preliminary technical evaluations of resolution and accuracy are also presented. The results of the evaluations reveal that the system statistically significantly improves surgical precision (p <; 0.01) with only a small increase in operating time. The force data recorded also reveals that operating force can be significantly affected by ergonomic factors, and that warnings for excessive force and workspace limits are needed.
Keywords
ergonomics; medical robotics; surgery; REMS; ergonomic factors; head surgery; microlaryngeal phonosurgery; microsurgical robotic system; neck surgery; preliminary evaluation; robotic ENT microsurgery system; Force; Force sensors; Joints; Neck; Robot sensing systems; Surgery;
fLanguage
English
Publisher
ieee
Conference_Titel
Intelligent Robots and Systems (IROS 2014), 2014 IEEE/RSJ International Conference on
Conference_Location
Chicago, IL
Type
conf
DOI
10.1109/IROS.2014.6942721
Filename
6942721
Link To Document