• DocumentCode
    137723
  • Title

    Preliminary evaluation of a new microsurgical robotic system for head and neck surgery

  • Author

    Olds, Kevin C. ; Chalasani, Preetham ; Pacheco-Lopez, Paulette ; Iordachita, Iulian ; Akst, Lee M. ; Taylor, Russell H.

  • Author_Institution
    Johns Hopkins Univ., Baltimore, MD, USA
  • fYear
    2014
  • fDate
    14-18 Sept. 2014
  • Firstpage
    1276
  • Lastpage
    1281
  • Abstract
    This paper presents an implementation and evaluation of the Robotic ENT Microsurgery System (REMS). The implementation is discussed in reference to analysis from previous work, and evaluated using a simulated surgical task designed to resemble microlaryngeal phonosurgery. Preliminary technical evaluations of resolution and accuracy are also presented. The results of the evaluations reveal that the system statistically significantly improves surgical precision (p <; 0.01) with only a small increase in operating time. The force data recorded also reveals that operating force can be significantly affected by ergonomic factors, and that warnings for excessive force and workspace limits are needed.
  • Keywords
    ergonomics; medical robotics; surgery; REMS; ergonomic factors; head surgery; microlaryngeal phonosurgery; microsurgical robotic system; neck surgery; preliminary evaluation; robotic ENT microsurgery system; Force; Force sensors; Joints; Neck; Robot sensing systems; Surgery;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Intelligent Robots and Systems (IROS 2014), 2014 IEEE/RSJ International Conference on
  • Conference_Location
    Chicago, IL
  • Type

    conf

  • DOI
    10.1109/IROS.2014.6942721
  • Filename
    6942721