• DocumentCode
    1381530
  • Title

    The Effect of Humidity on the Calibration of Precision Air Capacitors

  • Author

    Ford, L.H.

  • Volume
    96
  • Issue
    39
  • fYear
    1949
  • fDate
    1/1/1949 12:00:00 AM
  • Firstpage
    13
  • Lastpage
    16
  • Abstract
    The paper describes measurements, at a frequency of 1 000 c/s, of the capacitance changes in air capacitors due to moisture in the atmosphere. It is shown that these changes may be considerably in excess of those due to the alterations in the permittivity of the air. Over the range of relative humidity 30¿65% the total change in capacitance can be about 3 parts in 10 000 at room temperatures. The full advantage of the otherwise high stability of a precision air capacitor is lost when this effect is ignored.
  • fLanguage
    English
  • Journal_Title
    Proceedings of the IEE - Part III: Radio and Communication Engineering
  • Publisher
    iet
  • Type

    jour

  • DOI
    10.1049/pi-3.1949.0003
  • Filename
    5241089