The Comparison and Calibration of Power-Measuring Equipment at Wavelengths of 3 CM and 10 CM
Author
Bailey, R.A. ; French, H.A. ; Lane, J.A.
Volume
101
Issue
73
fYear
1954
fDate
9/1/1954 12:00:00 AM
Firstpage
325
Lastpage
329
Abstract
A description is given of recent experiments which have resulted in the production of substandard power-measuring equipment for wavelengths in the 3cm and 10cm wavebands. The equipment consists of two calibrated thermistor bridges measuring up to 1 mW of microwave power, and these may be regarded as the present United Kingdom substandards. The accuracy of the final calibration carried out at wavelengths of 3.18 and 10.77cm is estimated at ±2% and ±1% respectively. Good agreement was obtained in a comparison of the two standards used in the 3cm calibration, the water calorimeter and the torqueoperated vane wattmeter developed by Cullen and Stephenson.
Keywords
high-frequency measurement;