    The Comparison and Calibration of Power-Measuring Equipment at Wavelengths of 3 CM and 10 CM

  • Author

    Bailey, R.A. ; French, H.A. ; Lane, J.A.

  • Volume
    101
  • Issue
    73
  • fYear
    1954
  • fDate
    9/1/1954 12:00:00 AM
  • Firstpage
    325
  • Lastpage
    329
  • Abstract
    A description is given of recent experiments which have resulted in the production of substandard power-measuring equipment for wavelengths in the 3cm and 10cm wavebands. The equipment consists of two calibrated thermistor bridges measuring up to 1 mW of microwave power, and these may be regarded as the present United Kingdom substandards. The accuracy of the final calibration carried out at wavelengths of 3.18 and 10.77cm is estimated at ±2% and ±1% respectively. Good agreement was obtained in a comparison of the two standards used in the 3cm calibration, the water calorimeter and the torqueoperated vane wattmeter developed by Cullen and Stephenson.
  • Keywords
    high-frequency measurement;