DocumentCode :
1387457
Title :
A neuro-stimulus chip with telemetry unit for retinal prosthetic device
Author :
Liu, Wentai ; Vichienchom, Kasin ; Clements, Mark ; DeMarco, Stephen C. ; Hughes, Chris ; McGucken, Elliot ; Humayun, Mark S. ; De Juan, Eugene ; Weiland, James D. ; Greenberg, Robert
Author_Institution :
Dept. of Electr. & Comput. Eng., North Carolina State Univ., Raleigh, NC, USA
Volume :
35
Issue :
10
fYear :
2000
Firstpage :
1487
Lastpage :
1497
Abstract :
In this retinal prosthesis project, a rehabilitative device is designed to replace the functionality of defective photoreceptors in patients suffering from retinitis pigmentosa (RP) and age-related macular degeneration (AMD). The device consists of an extraocular and an intraocular unit. The implantable component receives power and a data signal via a telemetric inductive link between the two units. The extraocular unit includes a video camera and video processing board, a telemetry protocol encoder chip, and an RF amplifier and primary coil. The intraocular unit consists of a secondary coil, a rectifier and regulator, a retinal chip with a telemetry protocol decoder, a stimulus signal generator, and an electrode array. This paper focuses on the design, fabrication, and testing of a microchip which serves as the telemetry protocol decoder and stimulus signal generator. It is fabricated by MOSIS with 1.2-mm CMOS technology and was demonstrated to provide the desired biphasic current stimulus pulses for an array of 100 retinal electrodes at video frame rates.
Keywords :
CMOS integrated circuits; biomedical electronics; biomedical telemetry; eye; medical signal processing; neuromuscular stimulation; prosthetics; video cameras; video signal processing; 1.2 mm; CMOS technology; MOSIS; RF amplifier; age-related macular degeneration; biphasic current stimulus pulses; electrode array; extraocular unit; implantable component; intraocular unit; neuro-stimulus chip; primary coil; protocol encoder chip; retinal chip; retinal prosthetic device; retinitis pigmentosa; stimulus signal generator; telemetric inductive link; telemetry protocol; telemetry unit; video camera; video processing board; CMOS technology; Coils; Decoding; Electrodes; Photoreceptors; Prosthetics; Protocols; Retina; Signal generators; Telemetry;
fLanguage :
English
Journal_Title :
Solid-State Circuits, IEEE Journal of
Publisher :
ieee
ISSN :
0018-9200
Type :
jour
DOI :
10.1109/4.871327
Filename :
871327
Link To Document :
بازگشت