DocumentCode :
1387669
Title :
Time-Division SQUID Multiplexers With Reduced Sensitivity to External Magnetic Fields
Author :
Stiehl, G.M. ; Cho, H.-M. ; Hilton, G.C. ; Irwin, K.D. ; Mates, J.A.B. ; Reintsema, C.D. ; Zink, B.L.
Author_Institution :
Nat. Inst. of Stand. & Technol., Boulder, CO, USA
Volume :
21
Issue :
3
fYear :
2011
fDate :
6/1/2011 12:00:00 AM
Firstpage :
298
Lastpage :
301
Abstract :
Time-division SQUID multiplexers are used in many applications that require exquisite control of systematic error. One potential source of systematic error is the pickup of external magnetic fields in the multiplexer. We present measurements of the field sensitivity figure of merit, effective area, for both the first stage and second stage SQUID amplifiers in three NIST SQUID multiplexer designs. These designs include a new variety with improved gradiometry that significantly reduces the effective area of both the first and second stage SQUID amplifiers.
Keywords :
SQUIDs; magnetic fields; NIST SQUID multiplexer designs; SQUID amplifiers; external magnetic fields; field sensitivity; gradiometry; time-division SQUID multiplexers; Coils; Couplings; Magnetic separation; Multiplexing; NIST; SQUIDs; Superconducting magnets; Gradiometry; SQUID; multiplexer; transition Edge Sensor;
fLanguage :
English
Journal_Title :
Applied Superconductivity, IEEE Transactions on
Publisher :
ieee
ISSN :
1051-8223
Type :
jour
DOI :
10.1109/TASC.2010.2091483
Filename :
5643961
Link To Document :
بازگشت