DocumentCode
1388984
Title
Design of a CZT based breast SPECT system
Author
Singh, Manbir ; Mumcuoglu, Erkan
Author_Institution
Dept. of Radiol., Univ. of Southern California, Los Angeles, CA, USA
Volume
45
Issue
3
fYear
1998
fDate
6/1/1998 12:00:00 AM
Firstpage
1158
Lastpage
1165
Abstract
A high-resolution SPECT instrument dedicated to breast imaging has been designed incorporating arrays of collimated Cadmium-Zinc-Telluride (CZT) detectors tiled on either a cylindrical surface or a hemispherical surface surrounding the breast. The performance characteristics of a three-segment rotating parallel-hole collimator as well as a stationary multiple pin-hole collimator were considered for the cylindrical system. A stationary multiple pin-hole collimated system was also considered for the hemispherical design. Monte Carlo studies suggest that at almost equal spatial resolution of 0.5 cm, the cylindrical design with parallel-hole collimator would have an approximately a factor of two higher geometrical efficiency than the hemispherical pinhole collimated system including effects of attenuation in the breast. However, whereas the parallel-hole collimator must be rotated to acquire data from multiple angles of view, the pin-hole version has the advantage of recording data from 112 views in a stationary mode. Monte Carlo studies of filtered backprojection as well as a Bayesian reconstruction approach including attenuation and scatter within the breast, where the breast was modeled as a 15 cm hemisphere of uniform activity distribution containing three spherical lesions of diameters 1.0 cm, 0.7 cm and 0.5 cm respectively suggest that the 0.5 cm could be detected with either design in a one-hour SPECT study assuming a 10:1 tumor to background ratio. The authors conclude that a high resolution breast SPECT instrument where the resolution is limited to about 0.5 cm is viable with CZT detectors
Keywords
Bayes methods; Monte Carlo methods; biomedical equipment; image reconstruction; image resolution; single photon emission computed tomography; 0.5 to 15 cm; 1 h; CZT based breast SPECT system; Cd-Zn-Te; Monte Carlo studies; attenuation; breast imaging; collimated cadmium-zinc-telluride detectors; cylindrical surface; geometrical efficiency; hemispherical surface; medical diagnostic imaging; medical instrumentation; nuclear medicine; performance characteristics; scatter; spatial resolution; spherical lesions; stationary multiple pin-hole collimator; three-segment rotating parallel-hole collimator; Attenuation; Bayesian methods; Breast; Collimators; Detectors; High-resolution imaging; Instruments; Monte Carlo methods; Sensor arrays; Spatial resolution;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/23.681995
Filename
681995
Link To Document