DocumentCode
1389943
Title
Realisation of a small-size high resolution linear neutron scintillation detector
Author
Engels, R. ; Jansen, E. ; Reinartz, R. ; Reinhart, P. ; Schäfe, W. ; Schelten, J.
Author_Institution
Zentrallabor fur Electron., Forschungszentrum Julich GmbH, Germany
Volume
45
Issue
3
fYear
1998
fDate
6/1/1998 12:00:00 AM
Firstpage
502
Lastpage
504
Abstract
The spectrum of position sensitive neutron scintillation detectors, which have been developed and designed in our institute during the last decade, comprises several high resolution linear detectors. The design of the small size high resolution detector is based on a modified Anger technology using a linear array of 24 HAMAMATSU type R1770 rectangular photomultipliers and a 1 mm 6Li glass scintillator. The sensitive detector area is 200×20 mm and the spatial resolution is 1.2 mm. The neutron sensitivity at 1 Å is about 65% and the residual gamma sensitivity is less than 10-4 and the maximum count rate is about 100 kHz. The detector is linked to a highly flexible PC-based data acquisition system with 12 bit position and 16 bit time resolution. The stand alone detector and data acquisition system is aimed preferably at pulsed sources performing high resolution angle-dispersive time-of-flight experiments
Keywords
data acquisition; high energy physics instrumentation computing; microcomputer applications; neutron detection; nuclear electronics; photomultipliers; position sensitive particle detectors; solid scintillation detectors; 1 mm; 20 mm; 200 mm; 6Li glass scintillator; HAMAMATSU type R1770 rectangular photomultipliers; PC-based data acquisition; linear neutron scintillation detector; maximum count rate; modified Anger technology; neutron sensitivity; position sensitive neutron scintillation detectors; pulsed sources; residual gamma sensitivity; small-size high resolution scintillation detector; spatial resolution; Cathodes; Data acquisition; Glass; Instruments; Neutrons; Photomultipliers; Position sensitive particle detectors; Scintillation counters; Solid scintillation detectors; Spatial resolution;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/23.682435
Filename
682435
Link To Document