    In this study, the muscle fatigue detection capability of bipolar and unipolar lead systems used for surface electromyogram measurement was verified by simulation. The constructed model simplified the isometric contraction of the biceps brachii. There were two simulation experiments: 1) the addition and deletion of white noise and 2) the addition and deletion of hum noise. The pattern result of simulation 1) suggested the possibility that the muscle fatigue detection capability of a unipolar-leads system was high. The pattern 2) result showed the unipolar-leads system had a small influence of filtering, and suggested that the mixing of hum noise could be disregarded.
  • Keywords
    biomechanics; electromyography; fatigue; medical signal processing; muscle; neurophysiology; white noise; biceps brachii; bipolar lead systems; hum noise; isometric contraction; muscle fatigue detection verification; surface electromyogram measurement; surface electromyogram model; unipolar-lead system; white noise; Electrodes; Fatigue; Filtering; Muscles; Physiology; White noise;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society (EMBC), 2014 36th Annual International Conference of the IEEE
  • Conference_Location
    Chicago, IL
  • ISSN
    1557-170X
  • Type

    conf

  • DOI
    10.1109/EMBC.2014.6943541
  • Filename
    6943541