DocumentCode :
1390501
Title :
Effect of shielding the radioactivity outside the field of view on image quality in a dual head coincidence [PET camera]
Author :
Sossi, V. ; Pointon, B. ; Cohen, P. ; Johnson, R.R. ; Ruth, T.J.
Author_Institution :
TRIUMF, British Columbia Univ., Vancouver, BC, Canada
Volume :
47
Issue :
4
fYear :
2000
fDate :
8/1/2000 12:00:00 AM
Firstpage :
1561
Lastpage :
1566
Abstract :
Limited count rate ability and open geometry make the image quality obtained from dual head cameras very sensitive to radioactivity outside the camera field of view. The spatial distribution of non true events, scatter and randoms, originating from radioactivity outside the FOV was investigated. The image quality quantified with the noise equivalent count rate, improved by up to 58% as a consequence of shielding. Radioactivity in the first 10 cm outside the FOV was found to contribute most significantly to the detected events. A modular and easily movable patient body shield was designed
Keywords :
biomedical equipment; cameras; gamma-ray scattering; positron emission tomography; radioactivity; shielding; 10 cm; dual head coincidence PET camera; image quality; medical diagnostic imaging; medical instrumentation; modular easily movable patient body shield; noise equivalent count rate; nontrue events; nuclear medicine; radioactivity shielding effect; spatial distribution; Cameras; Degradation; Detectors; Event detection; Geometry; Image quality; Magnetic heads; Physics; Positron emission tomography; Scattering;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/23.873016
Filename :
873016
Link To Document :
بازگشت