Abstract :
A new calibration method related to a twin-type microcalorimeter is proposed for high-frequency power sensors. This method is based on intrinsic compensation of feeding line losses, operation that results in a unitary-gain microcalorimeter, whose mathematical model becomes similar to that of the lossless ideal system. Moreover, the time-consuming procedure required for a proper and autonomous microcalorimeter calibration, based on short or open-circuit load, is not necessary, and the microcalorimeter mount-dismount operation between raw measurements and microcalorimeter calibration steps disappears. The overall measurement time becomes about two times shorter when this system is used for the effective efficiency calibration of power sensors.
Keywords :
calibration; calorimeters; feeding line losses; high-frequency power sensors; intrinsic compensation; power sensor calibration; true-twin microcalorimeter; twin-type microcalorimeter; Calibration; Microwave measurements; Power measurement; Transmission line measurements; Correction factor; effective efficiency; insulated line; microcalorimeter; power sensor; switching method;
fLanguage :
English
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9456
Type :
jour
DOI :
10.1109/TIM.2010.2091300
Filename :
5658151
Link To Document :
بازگشت