Title :
Atomic Microwave Power Standard Based on the Rabi Frequency
Author :
Kinoshita, Moto ; Shimaoka, Kazuhiro ; Komiyama, Koji
Author_Institution :
Nat. Metrol. Inst. of Japan, Nat. Inst. of Adv. Ind. Sci. & Technol. (AIST), Tsukuba, Japan
fDate :
7/1/2011 12:00:00 AM
Abstract :
The Rabi frequency for an atomic hyperfine structure transition is proportional to the magnetic field strength of the incident electromagnetic wave. Using the known proportionality constant, an atomic microwave power standard based on the Rabi frequency can be realized. In this paper, the Rabi frequency of cesium atoms in a glass cell inserted in a microwave waveguide was measured using the atomic-candle method. The Rabi frequency was converted into the incident microwave power via the magnetic field strength by correcting the effects of the glass cell and impedance mismatch. The relative expanded uncertainty was approximately 4% at a coverage factor of 2. The microwave power measured from the Rabi frequency was compared with that measured by a calorimetric method. These results coincided within their uncertainties.
Keywords :
atomic clocks; hyperfine structure; magnetic field measurement; measurement standards; microwave measurement; Rabi frequency; atomic hyperfine structure transition; atomic microwave power standard; atomic-candle method; cesium atoms; glass cell; incident electromagnetic wave; magnetic field strength; microwave waveguide; Atomic clocks; Atomic measurements; Frequency measurement; Microwave measurements; Microwave theory and techniques; Power measurement; Uncertainty; Atom optics; atomic measurements; frequency measurement; measurement standards; metrology; microwave measurements; microwave sensors; power measurement;
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
DOI :
10.1109/TIM.2010.2091299