  • Abstract
    We conceived new oscillometric blood pressure (BP) estimation methods based on physical modeling. The crux of these methods is to simultaneously estimate the arterial stiffness and BP of the patient from a standard oscillometric waveform. Hence, in contrast to previous methods, the BP estimation is specific to the patient at the time of measurement and is robust against arterial stiffening. We conducted initial testing of one of the methods against invasive reference brachial BP measurements in eight cardiac catheterization patients before and after nitroglycerin infusions. The method achieved systolic, diastolic and mean BP root-mean-squared-errors of 7.2, 7.6 and 6.7 mmHg. These errors were, on average, 40% lower than an existing high-end method.
  • Keywords
    biomechanics; blood pressure measurement; blood vessels; arterial stiffness; brachial BP measurements; cardiac catheterization patients; model-based oscillometric blood pressure measurement; nitroglycerin infusions; oscillometric blood pressure estimation method; physical model; root-mean-squared-errors; standard oscillometric waveform; Biomedical monitoring; Blood pressure; Equations; Estimation; Mathematical model; Oscillators; Pressure measurement;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society (EMBC), 2014 36th Annual International Conference of the IEEE
  • Conference_Location
    Chicago, IL
  • ISSN
    1557-170X
  • Type

    conf

  • DOI
    10.1109/EMBC.2014.6943997
  • Filename
    6943997