    Electrical dental pulp testing. Defining parameters for proper instrumentation

  • Author

    Daskalov, I. ; Indjov, B. ; Mudrov, M.

  • Author_Institution
    Centre of Biomed. Eng., Bulgarian Acad. of Sci., Sofia, Bulgaria
  • Volume
    16
  • Issue
    1
  • fYear
    1997
  • Firstpage
    46
  • Lastpage
    50
  • Abstract
    The use of an electric pulp tester for quantitative measurement of the threshold excitation is justified only if its parameters meet the following specifications: rectangular stimulation pulse of 10 ms minimum duration; current-stabilized output with automatic increase of the intensity and peak current amplitude display; high output resistance of at least 500 M/spl Omega/; pulse repetition frequency of about 5 Hz at low intensities (up to 10-15 μA peak current intensity range 1 to 200-250 μA); active electrode contact area over 0.8 mm2 with application of electrolyte paste or use of conductive rubber tip; warning signal for limited current intensity; reference (return) electrode in the patient\´s hand or as a flat surface hook attached on the lip; both electrodes are sterilizable; pulp testers with intensity control and indication in arbitrary units, not calibrated in μA, are of poor diagnostic value and are not recommended. Devices with "yes" or "no" assessment of pulp excitability are also unacceptable. They do not use accurate current-stabilized generators and do not generate pulses with proper duration and shape.
  • Keywords
    bioelectric phenomena; biomedical equipment; test equipment; 1 to 250 muA; 10 ms; 15 Hz; 500 Mohm; conductive rubber tip; current-stabilized generators; diagnostic value; electrical dental pulp testing; electrolyte paste; flat surface hook; intensity control; limited current intensity; medical instrumentation; patient´s hand; proper instrumentation parameters definition; proper pulses generation; reference electrode; threshold excitation; warning signal; Area measurement; Automatic testing; Conductivity measurement; Current measurement; Dentistry; Electrical resistance measurement; Electrodes; Instruments; Pulse generation; Pulse measurements; Amplifiers; Bicuspid; Biomedical Engineering; Cuspid; Dental Enamel; Dental Pulp; Dental Pulp Test; Electric Impedance; Electric Stimulation; Electrodes; Electrophysiology; Equipment Design; Humans; Incisor; Molar; Reproducibility of Results; Signal Processing, Computer-Assisted; Stainless Steel; Surface Properties; Transistors;
  • fLanguage
    English
  • Journal_Title
    Engineering in Medicine and Biology Magazine, IEEE
  • Publisher
    ieee
  • ISSN
    0739-5175
  • Type

    jour

  • DOI
    10.1109/51.566152
  • Filename
    566152