• DocumentCode
    1401435
  • Title

    An absolute method for calibrating microwave attenuators

  • Author

    Davies, M.C.

  • Author_Institution
    General Electric Company Limited, Central Research Laboratories, Hirst Research Centre, Wembley, UK
  • Volume
    109
  • Issue
    23
  • fYear
    1962
  • fDate
    5/15/1905 12:00:00 AM
  • Firstpage
    796
  • Lastpage
    800
  • Abstract
    This method of calibrating a variable attenuator is based on the attenuation through a 3 dB coupler; an iterative process is used to take into account the absorption loss and the unequal power split of the coupler. Details are given of the procedure used for calibrating a particular attenuator, and the method of calibration at intervals of attenuation closer than the nominal 3 dB of the coupler is described. An estimate is made of the possible accuracy of the calibration, which, in the case described, was ±0.4 dB in 40 dB at a frequency of 6.2 Gc/s.
  • Keywords
    attenuation measurement;
  • fLanguage
    English
  • Journal_Title
    Proceedings of the IEE - Part B: Electronic and Communication Engineering
  • Publisher
    iet
  • ISSN
    0369-8890
  • Type

    jour

  • DOI
    10.1049/pi-b-2.1962.0135
  • Filename