Title :
Magnetic stimulation coil and circuit design
Author :
Davey, Kent ; Epstein, Charles M.
Author_Institution :
Neotonus Inc., New Smyrna Beach, FL, USA
Abstract :
A detailed analysis of the membrane voltage rise commensurate with the electrical charging circuit of a typical magnetic stimulator is presented. The analysis shows how the membrane voltage is linked to the energy, reluctance, and resonant frequency of the electrical charging circuit. There is an optimum resonant frequency for any nerve membrane depending on its capacitive time constant. The analysis also shows why a larger membrane voltage will be registered on the second phase of a biphasic pulse excitation. Typical constraints on three key quantities voltage, current, and silicone controlled rectifier (SCR) switching time dictate key components such as capacitance, inductance, and choice of turns.
Keywords :
bioelectric phenomena; biomagnetism; biomedical electronics; biomembranes; capacitance; circuit resonance; inductance; neurophysiology; switching; capacitive time constant; circuit design; electrical charging circuit; magnetic stimulation coil; membrane voltage rise; nerve membrane; optimum resonant frequency; resonant frequency; silicone controlled rectifier; switching time; Biomembranes; Circuit synthesis; Coils; Magnetic analysis; Magnetic circuits; Magnetic resonance; Magnetic stimulation; Resonant frequency; Thyristors; Voltage; Biomedical Engineering; Electric Stimulation Therapy; Equipment Design; Humans; Magnetics; Membrane Potentials; Nerve Fibers;
Journal_Title :
Biomedical Engineering, IEEE Transactions on