DocumentCode
1406233
Title
The radiation field characteristics of piezoelectric polymer membrane transducers when operating into air
Author
Hayward, Gordon ; Benny, Graham ; Banks, Robbie ; Galbraith, Walter
Author_Institution
Centre for Ultrasonic Eng., Strathclyde Univ., Glasgow, UK
Volume
47
Issue
6
fYear
2000
Firstpage
1438
Lastpage
1447
Abstract
This paper describes the evaluation of the radiation field characteristics of PVDF membrane transducers, whereby the active area is defined by an appropriate electrode pattern when operating into air. Measurement, including conventional scanning of the fields and laser vibrometry, is supplemented by finite element modeling over the frequency range 500 kHz to 1.1 MHz. Contrary to expectation, the surface displacements are nonuniform, giving rise to premature axial focusing of the field. This is shown to be caused by generation of the zero order symmetrical Lamb wave, arising from the electric field discontinuity at the electrode boundary of the transducer.
Keywords
finite element analysis; membranes; piezoelectric transducers; surface acoustic waves; ultrasonic transducers; 500 kHz to 1.1 MHz; PVDF membrane transducers; US transducers; active area; electric field discontinuity; electrode boundary; electrode pattern; finite element modeling; laser vibrometry; piezoelectric polymer membrane transducers; premature axial focusing; radiation field characteristics; surface displacements; zero order symmetrical Lamb wave; Biomedical transducers; Biomembranes; Ceramics; Electrodes; Frequency; Impedance; Manufacturing; Piezoelectric transducers; Polymers; Ultrasonic transducers;
fLanguage
English
Journal_Title
Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
Publisher
ieee
ISSN
0885-3010
Type
jour
DOI
10.1109/58.883533
Filename
883533
Link To Document