DocumentCode
1406303
Title
A 100-200 MHz ultrasound biomicroscope
Author
Knspik, D.A. ; Starkoski, Brian ; Pavlin, Charles J. ; Foste, F. Stuart
Author_Institution
Dept. of Med. Biophys., Toronto Univ., Ont., Canada
Volume
47
Issue
6
fYear
2000
Firstpage
1540
Lastpage
1549
Abstract
The development of higher frequency ultrasound imaging systems affords a unique opportunity to visualize living tissue at the microscopic level. This work was undertaken to assess the potential of ultrasound imaging in vivo using the 100-200 MHz range. Spherically focused lithium niobate transducers were fabricated. The properties of a 200 MHz center frequency device are described in detail. This transducer showed good sensitivity with an insertion loss of 18 dB at 200 MHz. Resolution of 14 /spl mu/m in the lateral direction and 12 /spl mu/m in the axial direction was achieved with f/1.14 focusing. A linear mechanical scan system and a scan converter were used to generate B-scan images at a frame rate up to 12 frames per second. System performance in B-mode imaging is limited by frequency dependent attenuation in tissues. An alternative technique, zone-focus image collection, was investigated to extend depth of field. Images of coronary arteries, the eye, and skin are presented along with some preliminary correlations with histology. These results demonstrate the feasibility of ultrasound biomicroscopy In the 100-200 MHz range. Further development of ultrasound backscatter imaging at frequencies up to and above 200 MHz will contribute valuable information about tissue microstructure.
Keywords
acoustic microscopes; backscatter; biomedical ultrasonics; cardiology; eye; skin; ultrasonic transducers; 100 to 200 MHz; 18 dB; B-scan images; LiNbO/sub 3/; coronary arteries; depth of field; eye; frequency dependent attenuation; histology; insertion loss; linear mechanical scan system; living tissue; scan converter; skin; spherically focused transducers; tissue microstructure; ultrasound backscatter imaging; ultrasound biomicroscope; ultrasound imaging systems; zone-focus image collection; Focusing; Frequency; High-resolution imaging; In vivo; Insertion loss; Lithium niobate; Microscopy; Ultrasonic imaging; Ultrasonic transducers; Visualization;
fLanguage
English
Journal_Title
Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
Publisher
ieee
ISSN
0885-3010
Type
jour
DOI
10.1109/58.883543
Filename
883543
Link To Document