Title :
Comparison of performances of electrical impedance tomography evaluated with 2-D and 3-D models
Author :
Chateaux, Jean-François ; Nadi, Mustapha
Author_Institution :
Fac. des Sci., Henri Poincare Univ., Paris, France
fDate :
11/1/2000 12:00:00 AM
Abstract :
This paper describes electrical impedance tomography (EIT) sensitivity evaluation using two models. The first is the classical circular two-dimensional (2-D) model used by past authors. The second is a three-dimensional (3-D) cylindrical model, which takes into account the height of the object under study. Having reported the analytical expression of the potential field of the 2-D model, the authors derive an equivalent solution for the 3-D case. Having analyzed the convergence of the solutions, they compute for different conductivity perturbations the ratio of the sensitivities obtained with the two models. Results indicate that the 2-D model, as compared with the 3-D model, generally overestimates sensitivity by a factor of two unless the conductivity perturbation is extensive and more conductive than the principal medium. In such a case, the 2-D model tends to underestimate EIT sensitivity.
Keywords :
electric impedance imaging; modelling; 2-D model; 3-D cylindrical model; conductivity perturbation; conductivity perturbations; electrical impedance tomography; medical diagnostic imaging; object height; Biomedical engineering; Biomedical imaging; Biomedical monitoring; Conductivity; Electrodes; Impedance measurement; Numerical models; Performance evaluation; Tomography; Two dimensional displays;
Journal_Title :
Microwave Theory and Techniques, IEEE Transactions on