DocumentCode
1406992
Title
Chronoamperometry Using Integrated Potentiostat Consisting of Poly-Si Thin-Film Transistors
Author
Kimura, Mutsumi ; Bundo, Kosuke ; Imuro, Yoshiki ; Sagawa, Yuki ; Setsu, Koushi
Volume
32
Issue
2
fYear
2011
Firstpage
212
Lastpage
214
Abstract
We have conducted chronoamperometry using an integrated potentiostat consisting of poly-Si thin-film transistors. An electrochemical measurement of enzymatic redox reaction using glucose oxidase and ferrocene mediator is attempted as a practical example. It is found that the chronoamperogram depends on the glucose concentration, and a glucose concentration as low as 2 mM can be detected by comparing the chronoamperogram without glucose. The glucose sensitivity of the chronoamperometry is similar to that of the cyclic voltammetry and sufficient for diabetic diagnosis.
Keywords
electrochemistry; enzymes; silicon; thin film transistors; voltammetry (chemical analysis); Si; chronoamperometry; cyclic voltammetry; diabetic diagnosis; electrochemical measurement; enzymatic redox reaction; ferrocene mediator; glucose concentration; glucose oxidase; glucose sensitivity; integrated potentiostat; poly-Si thin-film transistor; Chronoamperometry; poly-Si; potentiostat; thin-film transistor (TFT);
fLanguage
English
Journal_Title
Electron Device Letters, IEEE
Publisher
ieee
ISSN
0741-3106
Type
jour
DOI
10.1109/LED.2010.2090940
Filename
5671462
Link To Document