• DocumentCode
    1409370
  • Title

    Digital transfer voltmeters: principles and error characteristics

  • Author

    Deist, F. ; Kitai, R.

  • Author_Institution
    IBM Deutschland GmbH, WTEC, German Laboratories, West Germany
  • Volume
    110
  • Issue
    10
  • fYear
    1963
  • fDate
    10/1/1963 12:00:00 AM
  • Firstpage
    1887
  • Lastpage
    1904
  • Abstract
    Precision measurements of the r.m.s. value of a fluctuating voltage may be made using sampling and digital techniques. Although considerably more complex than analogue methods, the sampling voltmeter has substantial advantages in many respects. The paper discusses the principles on which the meter is based. Its error characteristics are derived and analysed for four periodic waveforms and for white noise. Using the results of the error analysis, design criteria are discussed. Finally, the principal requirements for a special-purpose digital computer to produce the meter reading are established.
  • Keywords
    computer applications; voltage measurement;
  • fLanguage
    English
  • Journal_Title
    Electrical Engineers, Proceedings of the Institution of
  • Publisher
    iet
  • ISSN
    0020-3270
  • Type

    jour

  • DOI
    10.1049/piee.1963.0267
  • Filename
    5247280