• DocumentCode
    141068
  • Title

    An evaluation of exhaled flow measuring mouthpieces for breath sampling devices

  • Author

    de Silva, Geethanga ; Beyette, Fred R.

  • Author_Institution
    Dept. of Electr. Eng. & Comput. Syst., Univ. of Cincinnati, Cincinnati, OH, USA
  • fYear
    2014
  • fDate
    26-30 Aug. 2014
  • Firstpage
    5768
  • Lastpage
    5771
  • Abstract
    A study is conducted to determine the dimensions of an exhaled flow measuring mouthpiece for a breath-sampling device that requires breathing for an extended period of time. Fleisch, Pitot and Venturi type differential pressure based flow measuring mouthpieces of various dimensions are evaluated. Inner diameters (IDs) of the cylinder shaped Fleisch, Pitot and Venturi mouthpieces are varied from 5mm to 25mm. Based on the study, we conclude IDs ranging from 8.75mm to 12.5mm for Fleisch type, IDs ranging from 10mm to 12.5mm for Pitot type and ID of 10mm for Venturi type are the most suitable dimensions for a breath sampling device.
  • Keywords
    biomedical equipment; biomedical measurement; pneumodynamics; Fleisch mouthpiece; Pitot mouthpiece; Venturi mouthpiece; breath sampling device; cylinder shaped mouthpiece; differential pressure mouthpiece; exhaled flow measuring mouthpiece; inner diameter; size 10 mm to 12.5 mm; size 5 mm to 25 mm; size 8.75 mm to 12.5 mm; Distance measurement; Electron tubes; Fluid flow measurement; Ports (Computers); Pressure measurement; Standards; Three-dimensional displays;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society (EMBC), 2014 36th Annual International Conference of the IEEE
  • Conference_Location
    Chicago, IL
  • ISSN
    1557-170X
  • Type

    conf

  • DOI
    10.1109/EMBC.2014.6944938
  • Filename
    6944938