Title :
A stereotaxic image-guided surgical robotic system for depth electrode insertion
Author :
Fanle Meng ; Hui Ding ; Guangzhi Wang
Author_Institution :
Dept. of Biomed. Eng., Tsinghua Univ., Beijing, China
Abstract :
This article constructs a surgical robotic system for the stereotactic insertion of the depth electrodes for stereoelectroencephalogram (SEEG). The purpose is to increase the efficiency of the stereotactic insertion of the electrodes. The registration method of this system is based on the noninvasive fiducial markers. After registration, the robotic system can locate all the preplanned electrode trajectories automatically. The validation of this proposed system has been performed by testing the time consumption of the system workflow and measuring the positioning accuracy on phantoms. From the result, we conclude that this surgical robotic system can assist surgeons in performing the stereotactic insertion of the depth electrodes accurately and efficiently.
Keywords :
biomedical electrodes; computerised tomography; electroencephalography; image registration; medical image processing; medical robotics; phantoms; position control; robot vision; surgery; SEEG; depth electrode insertion; noninvasive fiducial markers; phantoms; preplanned electrode trajectories; registration method; stereoelectroencephalogram; stereotactic insertion efficiency; stereotaxic image-guided surgical robotic system; Accuracy; Electrodes; Robot kinematics; Robot sensing systems; Surgery; Trajectory;
Conference_Titel :
Engineering in Medicine and Biology Society (EMBC), 2014 36th Annual International Conference of the IEEE
Conference_Location :
Chicago, IL
DOI :
10.1109/EMBC.2014.6945037