DocumentCode :
1413821
Title :
The System Design, Engineering Architecture, and Preliminary Results of a Lower-Cost High-Sensitivity High-Resolution Positron Emission Mammography Camera
Author :
Zhang, Yuxuan ; Ramirez, Rocio A. ; Li, Hongdi ; Liu, Shitao ; An, Shaohui ; Wang, Chao ; Baghaei, Hossain ; Wong, Wai-Hoi
Author_Institution :
M.D. Anderson Cancer Center, Univ. of Texas, Houston, TX, USA
Volume :
57
Issue :
1
fYear :
2010
Firstpage :
104
Lastpage :
110
Abstract :
A lower-cost high-sensitivity high-resolution positron emission mammography (PEM) camera is developed. It consists of two detector modules with the planar detector bank of 20 × 12 cm2. Each bank has 60 low-cost PMT-Quadrant-Sharing (PQS) LYSO blocks arranged in a 10 × 6 array with two types of geometries. One is the symmetric 19.36 × 19.36 mm2 block made of 1.5 × 1.5 × 10 mm3 crystals in a 12 × 12 array. The other is the 19.36 × 26.05 mm2 asymmetric block made of 1.5 × 1.9 × 10 mm3 crystals in 12 × 13 array. One row (10) of the elongated blocks are used along one side of the bank to reclaim the half empty PMT photocathode in the regular PQS design to reduce the dead area at the edge of the module. The bank has a high overall crystal packing fraction of 88%, which results in a very high sensitivity. Mechanical design and electronics have been developed for low-cost, compactness, and stability purposes. Each module has four Anger-HYPER decoding electronics that can handle a count-rate of 3 Mcps for single events. A simple two-module coincidence board with a hardware delay window for random coincidences has been developed with an adjustable window of 6 to 15 ns. Some of the performance parameters have been studied by preliminary tests and Monte Carlo simulations, including the crystal decoding map and the 17% energy resolution of the detectors, the point source sensitivity of 11.5% with 50 mm bank-to-bank distance, the 1.2 mm-spatial resolutions, 42 kcps peak Noise Equivalent Count Rate at 7.0-mCi total activity in human body, and the resolution phantom images. Those results show that the design goal of building a lower-cost, high-sensitivity, high-resolution PEM detector is achieved.
Keywords :
Monte Carlo methods; cancer; image resolution; mammography; particle detectors; positron emission tomography; Anger-HYPER decoding electronics; Monte Carlo simulations; PEM detector; PMT photocathode; breast cancer diagnosis; distance 50 mm; energy resolution; hardware delay window; high-resolution positron emission mammography camera; high-sensitivity positron emission mammography camera; human body; low-cost PMT-quadrant-sharing LYSO blocks; lower-cost positron emission mammography camera; mechanical design; noise equivalent count rate; point source sensitivity; random coincidences; resolution phantom images; spatial resolutions; two-module coincidence board; Cameras; Crystals; Decoding; Design engineering; Detectors; Energy resolution; Image resolution; Mammography; Radioactive decay; Systems engineering and theory; Nuclear imaging; PET instrumentation; nuclear medical applications;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.2009.2031644
Filename :
5409985