DocumentCode
1415126
Title
Digital Oscilloscope Calibration Using Asynchronously Sampled Signal Estimation
Author
Lapuh, R. ; Pinter, B. ; Voljc, B. ; Svetik, Z. ; Lindic, M.
Author_Institution
Metrol. Inst. of the Republic of Slovenia, Celje, Slovenia
Volume
60
Issue
7
fYear
2011
fDate
7/1/2011 12:00:00 AM
Firstpage
2570
Lastpage
2577
Abstract
Oscilloscopes are typically calibrated using special signals for each parameter. A common method using only sine-wave signals and analysis of sampled data is proposed to calibrate all major parameters of a digital sampling oscilloscope, greatly simplifying the calibration procedure while attaining lower uncertainties. The procedure can be also fully automated as the readout from the oscilloscope screen is no longer needed.
Keywords
calibration; digital storage oscilloscopes; signal sampling; asynchronously sampled signal estimation; digital oscilloscope calibration; sampled data analysis; sine-wave signal; Accuracy; Calibration; Frequency estimation; Harmonic distortion; Oscilloscopes; Uncertainty; Bandwidth; calibration; digital oscilloscope; frequency response; harmonic distortions; signal estimation;