Abstract
Sudomotor dysfunction is a major clinical manifestation of diabetic autonomic neuropathy and is an early sign of prediabetes. SUDOSCAN™, a noninvasive and quick method was developed to assess sudomotor function based on measurement of the electrical current response of the skin when different rectangular pulses of low voltage amplitudes are applied on nickel electrodes. Electrochemical skin conductance is then calculated from the resulting voltage and the generated current. This leads to an accurate noninvasive way to predict dysglycemia, including type II diabetes, without fasting requirement due to the sensitivity of Ni electrodes to human eccrine glands sweat.
Keywords
bioelectric potentials; biomedical electrodes; biomedical measurement; diseases; galvanometers; patient diagnosis; skin; SUDOSCAN; clinical manifestation; diabetes application; diabetic autonomic neuropathy; dysglycemia; early diagnosis; electrical current response measurement; electrochemical skin conductance; human eccrine glands sweat; nickel electrodes; noninvasive galvanic skin sensor; prediabetes; rectangular pulses; sudomotor dysfunction; type II diabetes; Current measurement; Diabetes; Electric potential; Electrodes; Foot; Nickel; Skin; Diabetes; diagnostic; electrochemistry; nickel electrode; reverse iontophoresis; skin conductance; sweat;
fLanguage
English
Journal_Title
Sensors Journal, IEEE
Publisher
ieee
ISSN
1530-437X
Type
jour
DOI
10.1109/JSEN.2010.2103308
Filename
5678618
Link To Document