    Parallax correction in PET using depth of interaction information

  • Author

    MacDonald, Lawrence R. ; Dahlbom, Magnus

  • Author_Institution
    Div. of Nucl. Med. & Biophys., California Univ., Los Angeles, CA, USA
  • Volume
    45
  • Issue
    4
  • fYear
    1998
  • fDate
    8/1/1998 12:00:00 AM
  • Firstpage
    2232
  • Lastpage
    2237
  • Abstract
    The authors investigate the restoration of radial resolution for transaxially off-center sources in a PET ring detector configuration using directly measured depth of interaction (DOI) information. Lutetium oxyorthosilicate (LSO) crystals are coupled to a photodetector along the radial dimension of the PET ring configuration. Four millimeter wide crystals are segmented into 4 mm and 10 mm lengths in the radial dimension and optically isolated so that scintillation light is localized to the radial extent of the segment, giving 4 mm and 10 mm DOI resolution. The authors also segmented 2 mm wide crystals into 10 mm lengths. This DOI technique is unambiguous but not practical for a multi-ring PET scanner. The objective here is to quantify the intrinsic radial resolution as a function of position within the FOV and DOI resolution. The degradation in radial resolution across 75% of the field of view was reduced by 45% (50%) with DOI resolution equal to one-half (one-third) the crystal length. Improved resolution via narrower crystals was overshadowed by DOI effects at ⩾50% off-center positions in 30 mm deep crystals with 10 mm of FOI resolution. Computer simulated images that incorporate 10 mm and 15 mm DOI in 30 mm crystals show improved resolution uniformity over a system without DOI
  • Keywords
    image resolution; medical image processing; photodetectors; positron emission tomography; 10 mm; 2 mm; 30 mm; 4 mm; PET parallax correction; PET ring detector configuration; computer simulated images; depth of interaction information; improved resolution uniformity; lutetium oxyorthosilicate crystals; medical diagnostic imaging; nuclear medicine; radial dimension; radial resolution restoration; scintillation light; segmented crystals; transaxially off-center sources; Computational modeling; Computer simulation; Couplings; Crystals; Degradation; Image resolution; Image restoration; Photodetectors; Positron emission tomography; Solid scintillation detectors;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/23.708354
  • Filename
    708354