• DocumentCode
    1419731
  • Title

    Quantification of the 3-D electromagnetic power absorption rate in tissue during transurethral prostatic microwave thermotherapy using heat transfer model

  • Author

    Zhu, Liang ; Xu, Lisa X. ; Chencinski, Norbert

  • Author_Institution
    Dept. of Appl. Sci., City Univ. of New York, NY, USA
  • Volume
    45
  • Issue
    9
  • fYear
    1998
  • Firstpage
    1163
  • Lastpage
    1172
  • Abstract
    Experiments were performed in a tissue microwave-equivalent phantom gel to quantitatively examine the volumetric heating produced by a microwave antenna with a peripheral cooling system for the transurethral prostatic thermotherapy. Based on previous research, expression for the specific absorption rate (SAR) of microwave energy in the gel was extended to three dimensions, which includes its dependence on radial, angular, and axial direction. A theoretical heat transfer model was developed to study the temperature distribution in the gel by introducing this proposed SAR expression. The parameters in this expression and the convection coefficient due to the chilled water running around the antenna were determined using a least-square residual fit of the theoretical temperature predictions to the experimentally measured steady-state temperature field within the gel. The analytical expression of the three-dimensional SAR distribution obtained in this study will help provide a better understanding of the microwave heating pattern in the prostatic tissue and, thus, to aid in designing improved applicators. It can also be used in the future as an accurate input to heat transfer models which predict temperature distributions during the transurethral microwave thermotherapy.
  • Keywords
    microwave heating; physiological models; radiation therapy; temperature distribution; 3-D electromagnetic power absorption rate quantification; chilled water; experimentally measured steady-state temperature field; improved applicators design; least-square residual fit; microwave antenna; peripheral cooling system; specific absorption rate; theoretical heat transfer model; theoretical temperature predictions; tissue microwave-equivalent phantom gel; transurethral prostatic microwave thermotherapy; volumetric heating; Electromagnetic heating; Electromagnetic modeling; Electromagnetic wave absorption; Heat transfer; Imaging phantoms; Medical treatment; Microwave technology; Power system modeling; Temperature distribution; Temperature measurement; Gels; Humans; Hyperthermia, Induced; Least-Squares Analysis; Male; Microwaves; Models, Biological; Prostatic Hyperplasia; Temperature; Thermal Conductivity; Urethra;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/10.709560
  • Filename