• DocumentCode
    1422173
  • Title

    Ensemble tracking for 2D vector velocity measurement: Experimental and initial clinical results

  • Author

    Bohs, Laurence N. ; Geiman, Beth J. ; Anderson, Martin E. ; Breit, Sean M. ; Trahey, Gregg E.

  • Author_Institution
    Dept. of Biomed. Eng., Duke Univ., Durham, NC, USA
  • Volume
    45
  • Issue
    4
  • fYear
    1998
  • fDate
    7/1/1998 12:00:00 AM
  • Firstpage
    912
  • Lastpage
    924
  • Abstract
    We describe a new method, called ensemble tracking, for estimating two-dimensional velocities with ultrasound. Compared to previous speckle tracking techniques, ensemble tracking measures motion over smaller times and distances, increasing maximum velocities and reducing errors due to echo decorrelation. Ensemble tracking uses parallel receive processing, 2D pattern matching, and interpolation of the resulting tracking grid to estimate sub-pixel speckle translations between successive ultrasonic acquisitions. In this study, small translations of a tissue mimicking phantom were quantified at transducer angles of 0/spl deg/, 45/spl deg/, and 90/spl deg/. Measurements over three parallel beam spacings and all transducer angles had mean errors from -4% to +11%, when parallel beam amplitudes were normalized. Such amplitude normalization substantially improved results at 45/spl deg/ and 90/spl deg/. The amplitude, spacing, and correlation between the parallel beams were quantified, and their effects on the accuracy and precision of estimates are discussed. Finally, initial clinical results demonstrate the ability to track and display blood flow in the carotid artery.
  • Keywords
    biomedical ultrasonics; blood flow measurement; tracking; ultrasonic velocity measurement; 2D pattern matching; 2D vector US velocity measurement; blood flow; carotid artery; ensemble tracking; interpolation; parallel beam spacings; parallel receive processing; sub-pixel speckle translations; tissue mimicking phantom; tracking grid; transducer angles; Decorrelation; Interpolation; Motion measurement; Pattern matching; Speckle; Tracking; Ultrasonic imaging; Ultrasonic transducers; Ultrasonic variables measurement; Velocity measurement;
  • fLanguage
    English
  • Journal_Title
    Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0885-3010
  • Type

    jour

  • DOI
    10.1109/58.710557
  • Filename
    710557