• DocumentCode
    1422767
  • Title

    Design considerations in a myoelectric hand prosthesis

  • Author

    Fyson, J. ; Jilbert, P.H. ; Truscott, J.R.

  • Author_Institution
    United Kingdom Atomic Energy Authority, Atomic Weapons Research Establishment, Reading, UK
  • Volume
    116
  • Issue
    2
  • fYear
    1969
  • fDate
    2/1/1969 12:00:00 AM
  • Firstpage
    281
  • Lastpage
    290
  • Abstract
    The design of a myoelectric control system, for use with an electrically powered hand prosthesis, is described. This system, developed for the Ministry of Health from an experimental system due to Bottomley and Cowell, has proportional control of velocity and force. Although functionally simple, the design is complicated by the need to fulfil a number of requirements, notably low power consumption, convenient size and weight, good reliability, ease of maintenance and ease of adjustment. A significant degree of power economy is achieved by automatic removal of the drive when the desired grip is attained. Important features of the design are given, and an analysis is made of the feedback system.
  • fLanguage
    English
  • Journal_Title
    Electrical Engineers, Proceedings of the Institution of
  • Publisher
    iet
  • ISSN
    0020-3270
  • Type

    jour

  • DOI
    10.1049/piee.1969.0052
  • Filename
    5249457