  • Abstract
    Defibrillators often use an ac measurement to ensure safe electrode placement. Some defibrillators also utilize impedance measurements for ventilation monitoring. This paper investigates how such measurements can be optimized for high respiration sensitivity and finds that it is beneficial to add two extra electrodes in addition to the two defibrillator electrodes. This four electrode setup increases sensitivity and also allows respiration measurements at lower frequencies than the two electrode setup.
  • Keywords
    bioelectric phenomena; biomedical electrodes; defibrillators; electric impedance imaging; pneumodynamics; ac measurement; bioimpedance-based respiration monitoring; defibrillator; extra electrodes; high respiration sensitivity; respiration measurements; safe electrode placement; ventilation monitoring; Biomedical measurement; Electrical resistance measurement; Electrodes; Frequency measurement; Impedance; Impedance measurement; Ventilation; Bioimpedance; defibrillator; respiration; sensitivity; ventilation;