A New Indirect Calibration Method for an Equivalent Source Mismatch of a Power Splitter Using Measurements at Only One Port
Author :
Suzuki, Kiyoshi
Author_Institution :
Meas. Stand. Center, Agilent Technol. Int. Japan, Ltd., Tokyo, Japan
Volume :
60
Issue :
7
fYear :
2011
fDate :
7/1/2011 12:00:00 AM
Firstpage :
2684
Lastpage :
2689
Abstract :
A new measurement procedure is described for determining the equivalent source mismatch ΓG of a power splitter. It uses a vector network analyzer (VNA) configured for one-port calibration and measurement. The power splitter ports that are not connected to the VNA test port are loaded in turn with characterized standards, such as an open, a short, a load, and a termination. It also takes advantage of the reciprocity characteristic of the power splitter. As the system impedance is normalized to the impedance of the termination, the termination works as a perfect load in measurements and calculations for the S-parameters of the power splitter. ΓG is calculated from the renormalized S-parameters to Z0 . The propagation of uncertainties is analyzed.
Keywords :
S-parameters; calibration; electric sensing devices; network analysers; power measurement; S-parameter; equivalent source mismatch; indirect calibration method; power measurement; power sensor; power splitter; vector network analyzer; Calibration; Loading; Power measurement; Scattering parameters; Sensitivity; Standards; Uncertainty; Calibration; equivalent source mismatch; power sensor; power splitter; uncertainty;
fLanguage :
English
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9456
Type :
jour
DOI :
10.1109/TIM.2010.2096891
Filename :
5688454
Link To Document :
بازگشت