DocumentCode :
1428414
Title :
RF-DC differences of coaxial thermal standards
Author :
Huang, De-Xiang ; Chen, Ming-lu ; He, Shui-zhi
Author_Institution :
Nat. Inst. of Stand. & Technol., Gaithersburg, MD, USA
Volume :
39
Issue :
2
fYear :
1990
fDate :
4/1/1990 12:00:00 AM
Firstpage :
313
Lastpage :
317
Abstract :
The RF-DC differences (δ) of thermal converters are calculated from complex transmission line theory. The Boella effect and the skin effect are considered. The formula δ=Af1/2 +Bf2 can be obtained by experiments and proved by comparison of converters. The uncertainty of δ is 0.01% at 1 MHz and 0.1% at 30 MHz
Keywords :
measurement standards; transmission line theory; voltage measurement; 1 MHz; 30 MHz; Boella effect; RF-DC differences; coaxial thermal standards; skin effect; thermal converters; transmission line theory; voltage standard; Coaxial components; Estimation theory; Helium; Radio frequency; Resistors; Skin effect; Transmission line measurements; Transmission line theory; Uncertainty; Voltage;