RF heating of implanted spinal fusion stimulator during magnetic resonance imaging
Author :
Chou, Chung-Kwang ; McDougall, John A. ; Chan, Kwok W.
Author_Institution :
Dept. of Radiat. Res., City of Hope Nat. Med. Center, Duarte, CA, USA
Volume :
44
Issue :
5
fYear :
1997
fDate :
5/1/1997 12:00:00 AM
Firstpage :
367
Lastpage :
373
Abstract :
Radio frequency (RF) heating of an implanted spinal fusion stimulator (SpF) during magnetic resonance imaging (MRI) was studied on a full-size human phantom. Heating during MRI scans (GE Signa 4X, 1.5 T) was measured with RF-transparent fiberoptic sensors. With the implant correctly connected, the maximum temperature rises were less than 2°C during the 26 min that the scans were at maximum RF power. At the tip of a broken stimulator lead (connecting the SpF generator and its electrodes), the maximum temperature rise was 11-14°C. Regular 4-min scans of the spinal cord produced similar temperature rises at the broken tip. After the generator and the leads were removed, heating at the electrode connector tip was less than 1.5°C. The control temperature rises at the same locations, without the stimulator, were less than 0.5°C. This study shows that spinal fusion stimulator heating is within the Food and Drug Administration safety guideline of 2°C. However, if a lead wire is broken, it is unsafe during MRI scans. Radiological examinations will be necessary to ensure the integrity of the implant.
Keywords :
biological effects of radiation; biomechanics; biomedical NMR; bone; fibre optic sensors; health hazards; hyperthermia; patient treatment; radiofrequency heating; 1.5 T; 26 min; 4 min; Food and Drug Administration safety guideline; MRI scans; RF heating; RF-transparent fiberoptic sensors; bone fusion; broken stimulator lead; control temperature rises; electrode connector tip; full-size human phantom; implanted spinal fusion stimulator; magnetic resonance imaging; spinal cord; spinal fusion stimulator; temperature rises; Electrodes; Heating; Humans; Imaging phantoms; Implants; Magnetic resonance imaging; Magnetic sensors; Optical fiber sensors; Radio frequency; Temperature sensors; Electric Stimulation; Electrodes; Equipment Failure; Fiber Optics; Heat; Humans; Magnetic Resonance Imaging; Microwaves; Models, Anatomic; Phantoms, Imaging; Prostheses and Implants; Spinal Fusion; Surface Properties;
fLanguage :
English
Journal_Title :
Biomedical Engineering, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9294
Type :
jour
DOI :
10.1109/10.568912
Filename :
568912
Link To Document :
بازگشت