• DocumentCode
    1431286
  • Title

    Improving thermal ablation delineation with electrode vibration elastography using a bidirectional wave propagation assumption

  • Author

    DeWall, R.J. ; Varghese, Tomy

  • Author_Institution
    Depts. of Med. Phys. & Biomed. Eng., Univ. of Wisconsin-Madison, Madison, WI, USA
  • Volume
    59
  • Issue
    1
  • fYear
    2012
  • fDate
    1/1/2012 12:00:00 AM
  • Firstpage
    168
  • Lastpage
    173
  • Abstract
    Thermal ablation procedures are commonly used to treat hepatic cancers and accurate ablation representation on shear wave velocity images is crucial to ensure complete treatment of the malignant target. Electrode vibration elastography is a shear wave imaging technique recently developed to monitor thermal ablation extent during treatment procedures. Previous work has shown good lateral boundary delineation of ablated volumes, but axial delineation was more ambiguous, which may have resulted from the assumption of lateral shear wave propagation. In this work, we assume both lateral and axial wave propagation and compare wave velocity images to those assuming only lateral shear wave propagation in finite element simulations, tissue-mimicking phantoms, and bovine liver tissue. Our results show that assuming bidirectional wave propagation minimizes artifacts above and below ablated volumes, yielding a more accurate representation of the ablated region on shear wave velocity images. Area overestimation was reduced from 13.4% to 3.6% in a stiff-inclusion tissue-mimicking phantom and from 9.1% to 0.8% in a radio-frequency ablation in bovine liver tissue. More accurate ablation representation during ablation procedures increases the likelihood of complete treatment of the malignant target, decreasing tumor recurrence.
  • Keywords
    biomechanics; biomedical ultrasonics; biothermics; cancer; elastic waves; finite element analysis; image representation; liver; medical image processing; phantoms; axial delineation; bidirectional wave propagation assumption; bovine liver tissue; electrode vibration elastography; finite element simulation; hepatic cancers; image representation; lateral boundary delineation; lateral shear wave propagation; shear wave imaging; shear wave velocity images; thermal ablation delineation; tissue mimicking phantom; Electrodes; Finite element methods; Pathology; Phantoms; Propagation; Radio frequency; Vibrations; Animals; Artifacts; Cattle; Computer Simulation; Elasticity Imaging Techniques; Finite Element Analysis; Image Processing, Computer-Assisted; Laser Therapy; Liver; Phantoms, Imaging; Signal Processing, Computer-Assisted; Surgery, Computer-Assisted; Vibration;
  • fLanguage
    English
  • Journal_Title
    Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0885-3010
  • Type

    jour

  • DOI
    10.1109/TUFFC.2012.2169
  • Filename
    6138740