• DocumentCode
    1434769
  • Title

    A.C. bridge for platinum-resistance thermometry

  • Author

    Thompson, A.M. ; Small, G.W.

  • Author_Institution
    Commonwealth Scientific and Industrial Research Organisation (CSIRO), National Standards Laboratory, Division of Applied Physics, Sydney, Australia
  • Volume
    118
  • Issue
    11
  • fYear
    1971
  • fDate
    11/1/1971 12:00:00 AM
  • Firstpage
    1662
  • Lastpage
    1666
  • Abstract
    An a.c. bridge has been constructed for the measurement of the resistance of platinum-resistance thermometers. It is based on a multidial 2-stage ratio transformer, using a single reference resistor of 25¿, and operates at a frequency of 320Hz. The measuring current of 1 or ¿2 mA is carried by the standard and unknown resistors in series, and the potential drops are compared by means of the ratio transformer, using a single gain-stabilised detector. The ratio transformer is separately excited to ensure that the current in the potential circuit of the bridge is not significant. The bridge has five switched decades for a range of 111 ¿, and the residual unbalance is indicated on a direct reading meter with a most sensitive range of 1 m¿ full scale. Feedback of the quadrature component of signal from the detector eleminates the need for manual reactance balance. The resolution is better than 10 ¿¿ and the overall linearity better than 20 ¿¿. A recorder output (± 1V) is provided.
  • Keywords
    bridge instruments; electric sensing devices; platinum; resistance measurement; resistance thermometers; temperature measurement; thermometers; AC; bridge instruments; platinum resistance thermometer; resistance measurement; use of multidial two stage ratio transformer;
  • fLanguage
    English
  • Journal_Title
    Electrical Engineers, Proceedings of the Institution of
  • Publisher
    iet
  • ISSN
    0020-3270
  • Type

    jour

  • DOI
    10.1049/piee.1971.0306
  • Filename
    5251867