DocumentCode :
1437389
Title :
The measurement of absolute absorption of millimeter radiation in gases: the absorption of Co and O/sub 2/
Author :
Read, William G. ; Hillig, Kurt W., II ; Cohen, Edward A. ; Pickett, Herbert M.
Author_Institution :
Jet Propusion Lab., Pasadena, CA, USA
Volume :
36
Issue :
8
fYear :
1988
Firstpage :
1136
Lastpage :
1143
Abstract :
An apparatus is described that will measure absolute absorption of millimeter radiation in gases. The method measures the change in the quality factor of a Fabry-Perot resonator with and without gas present. The magnitude of the change is interpreted in terms of the absorption of the lossy medium inside the resonator. Experiments have been performed on the 115 GHz CO line and the 119 GHz O/sub 2/ line at two different temperatures to determine the linewidth parameter and the peak absorption value. These numbers can be combined to give the integrated intensity which can be accurately calculated from results of spectroscopy measurements. The CO results are within 2% percent of theoretically predicted valves. Measurements on O/sub 2/ have shown that absorption can be measured as 0.5 dB/km with this technique. Results have been obtained for oxygen absolute absorption in the 60-80 GHz region.<>
Keywords :
carbon compounds; electromagnetic wave absorption; microwaves; oxygen; 115 GHz; 119 GHz; 60 to 80 GHz; Co; Fabry-Perot resonator; O/sub 2/; absolute absorption; gases; integrated intensity; linewidth parameter; millimeter radiation; peak absorption value; quality factor; Electromagnetic wave absorption; Extraterrestrial measurements; Fabry-Perot; Gases; Geophysical measurements; Laboratories; Mirrors; Optical resonators; Propulsion; Spectroscopy;
fLanguage :
English
Journal_Title :
Antennas and Propagation, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-926X
Type :
jour
DOI :
10.1109/8.7226
Filename :
7226
Link To Document :