DocumentCode
1441040
Title
Time-scale analysis of quadrature Doppler ultrasound signals
Author
Aydin, N. ; Markus, H.S.
Author_Institution
Div. of Clinical Neurosci., St. George´´s Hospital Med. Sch., London, UK
Volume
148
Issue
1
fYear
2001
fDate
1/1/2001 12:00:00 AM
Firstpage
15
Lastpage
22
Abstract
Most Doppler ultrasound systems employ quadrature demodulation techniques at the detection stage. The information concerning flow direction, encoded in the phase relationship between in-phase and quadrature-phase channels, is not obvious at this stage. The complex fast Fourier transform can be used to obtain directional information in the frequency domain, as well as time frequency analysis of Doppler signals. However, it has an inherent time-frequency resolution limitation. The wavelet transform allows the time-frequency resolution compromise to be optimised. Mapping directional information in the scale domain is also desirable. A method is described, based on the utilisation of complex wavelets and negative scales. It eliminates the intermediate processing stages for obtaining directional Doppler signals for time-scale analysis
Keywords
Doppler effect; acoustic signal processing; demodulation; time-frequency analysis; ultrasonics; wavelet transforms; Doppler signals; complex fast Fourier transform; detection stage; directional information; flow direction; in-phase channels; negative scale; quadrature Doppler ultrasound signals; quadrature demodulation; quadrature-phase channels; time frequency analysis; time-frequency resolution; time-frequency resolution limitation; time-scale analysis; wavelet transform; wavelets;
fLanguage
English
Journal_Title
Science, Measurement and Technology, IEE Proceedings -
Publisher
iet
ISSN
1350-2344
Type
jour
DOI
10.1049/ip-smt:20010106