• DocumentCode
    1442354
  • Title

    Gamma-radiation imaging system based on the Compton effect

  • Author

    Everett, D.B. ; Fleming, J.S. ; Todd, R.W. ; Nightingale, J.M.

  • Author_Institution
    National Institution for Medical Research, London, UK
  • Volume
    124
  • Issue
    11
  • fYear
    1977
  • fDate
    11/1/1977 12:00:00 AM
  • Firstpage
    995
  • Abstract
    A novel gamma-ray camera is proposed that is based on the Compton effect. The basis of the device is a segmented semiconductor detector which is able to locate successive interaction points in multicollision trajectories and the associated energy losses. In this way, conic section can be computed on an image plane for each emission from a point source in the object. An overall image can be constructed by summation of the individual ellipses. The effect of measurement errors is investigated theoretically and by simulation, and leads to favourable estimates for the resolution and sensitivity of the camera. The question of image processing to improve performance is briefly discussed.
  • Keywords
    Compton effect; cameras; gamma-ray detection and measurement; image sensors; picture processing; scanning radiography; Compton effect; camera sensitivity; energy losses; gamma radiation imaging system; gamma ray camera; image processing; individual ellipses; measurement errors effect; multicollision trajectories; semiconductor detector; simulation;
  • fLanguage
    English
  • Journal_Title
    Electrical Engineers, Proceedings of the Institution of
  • Publisher
    iet
  • ISSN
    0020-3270
  • Type

    jour

  • DOI
    10.1049/piee.1977.0203
  • Filename
    5253391