• DocumentCode
    1452328
  • Title

    Validation of object-induced MR distortion correction for frameless stereotactic neurosurgery

  • Author

    Dean, David ; Kamath, Janardhan ; Duerk, Jeffrey L. ; Ganz, Edward

  • Author_Institution
    Dept. of Neurological Surg., Case Western Reserve Univ., Cleveland, OH, USA
  • Volume
    17
  • Issue
    5
  • fYear
    1998
  • Firstpage
    810
  • Lastpage
    816
  • Abstract
    Spatial fidelity is a paramount issue in image guided neurosurgery. Until recently, three-dimensional computed tomography (3D CT) has been the primary modality because it provides fast volume capture with pixel level (1 mm) accuracy. While three-dimensional magnetic resonance (3D MR) images provide superior anatomic information, published image capture protocols are time consuming and result in scanner- and object-induced magnetic field inhomogeneities which raise inaccuracy above pixel size. Using available scanner calibration software, a volumetric algorithm to correct for object-based geometric distortion, and a Fast Low Angle SHot (FLASH) 3D MR-scan protocol, the authors were able to reduce mean CT to MR skin-adhesed fiducial marker registration error from 1.36 to 1.09 mm. After dropping the worst one or two of six fiducial markers, mean registration error dropped to 0.62 mm (subpixel accuracy). Three dimensional object-induced error maps present highest 3D MR spatial infidelity at the tissue interfaces (skin/air, scalp/skull) where frameless stereotactic fiducial markers are commonly applied. The algorithm produced similar results in two patient 3D MR-scans.
  • Keywords
    biomedical MRI; brain; computerised tomography; image registration; magnetic susceptibility; medical image processing; surgery; CT to MR skin-adhesed fiducial marker registration error; frameless stereotactic neurosurgery; magnetic field inhomogeneities; magnetic resonance imaging; medical diagnostic imaging; object-based geometric distortion; object-induced MR distortion correction validation; pixel level accuracy; pixel size; published image capture protocols; spatial fidelity; three dimensional object-induced error maps; volumetric algorithm; Calibration; Computed tomography; Error correction; Magnetic fields; Magnetic resonance; Neurosurgery; Pixel; Protocols; Skin; Software algorithms; Algorithms; Head; Humans; Image Processing, Computer-Assisted; Magnetic Resonance Imaging; Neurosurgical Procedures; Stereotaxic Techniques;
  • fLanguage
    English
  • Journal_Title
    Medical Imaging, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0278-0062
  • Type

    jour

  • DOI
    10.1109/42.736049
  • Filename
    736049