DocumentCode
1453318
Title
Analytical calculation of photon distributions in SPECT projections
Author
Wells, R.G. ; Celler, A. ; Harrop, R.
Author_Institution
Dept. of Geophys. & Astron., British Columbia Univ., Vancouver, BC, Canada
Volume
45
Issue
6
fYear
1998
fDate
12/1/1998 12:00:00 AM
Firstpage
3202
Lastpage
3214
Abstract
In this paper the authors present a method for analytically calculating the distribution of photons detected in single photon emission computed tomography (SPECT) projections. The technique is applicable to sources in homogeneous and nonhomogeneous media. The photon distribution (primary, first-, and second-order Compton scatter) is computed using a precalculated camera-dependent look-up table in conjunction with an attenuation map of the scattering object and a map of the activity distribution. The speed and accuracy of this technique is compared to that of Monte Carlo simulations. The cases considered are a point source in a homogeneous and also in a nonhomogeneous scattering medium, an extended source in a nonhomogeneous medium, and a homogeneous cylinder filled uniformly with activity. The method is quantitatively accurate and faithfully reproduces the spatial distribution of the unscattered and scattered photons. For comparable statistical precision in the peak of the calculated distribution, their approach can result in a gain in calculation time over Monte Carlo simulators. For point sources, the computation times are improved by a factor of 20-150. However, this gain depends on the source configuration, and calculation times become comparable for an 800 voxel source and are five times slower for a 55000 voxel source. The method also offers an increase in the speed of computation of higher order Compton scatter events over a similar analytical technique
Keywords
medical image processing; single photon emission computed tomography; Monte Carlo simulations; SPECT projections; activity distribution; analytical calculation; higher order Compton scatter events; nonhomogeneous scattering medium; photon distribution; photon distributions; precalculated camera-dependent look-up table; single photon emission computed tomography; source configuration; Attenuation; Computational modeling; Distributed computing; Electromagnetic scattering; Monte Carlo methods; Nonhomogeneous media; Optical computing; Particle scattering; Single photon emission computed tomography; Table lookup;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/23.736199