• DocumentCode
    1455507
  • Title

    Ten-volt Josephson voltage standard at the ETL

  • Author

    Murayama, Yasushi ; Sakuraba, Toshiaki ; Sakamoto, Yasuhiko ; Iwasa, Akio ; Yoshida, Haruo ; Kozakai, Tsuneoki ; Endo, Tadashi

  • Author_Institution
    Electrotech. Lab., Ibaraki, Japan
  • Volume
    46
  • Issue
    2
  • fYear
    1997
  • fDate
    4/1/1997 12:00:00 AM
  • Firstpage
    233
  • Lastpage
    236
  • Abstract
    A voltage standard system using a 10 V Josephson junction array has been developed. The uncertainty is 0.6×10-8 for measurement of the 10 V output of a Zener diode reference standard. A comparison of the 10 V array system with the conventional 10 V measurement system composed of a 1 V array and a voltage divider was performed through measurements of a 10 V output of a Zener. The result agreed within the uncertainty (2×10-8) of the comparison. The 10 V output voltages of a Zener reference standard are measured consecutively by the 10 V array system
  • Keywords
    calibration; measurement errors; measurement standards; superconductor-insulator-superconductor devices; voltage measurement; 10 V; ETL; Electrotechnical Laboratory; Josephson junction array; Josephson voltage standard; Nb-AlO-Nb; Nb/AlOx/Nb junctions; Zener diode reference standard; uncertainty; voltage standard system; Calibration; Diodes; Josephson junctions; Laboratories; Millimeter wave measurements; Millimeter wave technology; Pollution measurement; Semiconductor device measurement; Standards development; Voltage;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/19.571820
  • Filename
    571820