DocumentCode :
1455603
Title :
Step number determination with two sets of frequency and step number in Josephson-junction array voltage standard
Author :
Ogita, Eiji ; Sakuma, Hiroaki ; Murayama, Yasushi ; Sakamoto, Yasuhiko
Author_Institution :
Yokogawa Electr. Corp., Yamanashi, Japan
Volume :
46
Issue :
2
fYear :
1997
fDate :
4/1/1997 12:00:00 AM
Firstpage :
253
Lastpage :
255
Abstract :
A method to determine the step number of quantized voltage steps in a Josephson-junction array by using two sets of frequency and step numbers is studied and an uncertainty analysis is given. The method can correctly determine the step number even with an initial voltage measurement that is less accurate (one or two orders) than that required for the conventional method
Keywords :
measurement errors; measurement standards; superconducting junction devices; voltage measurement; 10 V; Josephson-junction array voltage standard; i-V measurement; initial voltage measurement; quantized voltage steps; step number; uncertainty analysis; Equations; Fluctuations; Frequency; Helium; Measurement standards; Millimeter wave measurements; Millimeter wave technology; Uncertainty; Voltage measurement; Voltmeters;