Title :
AC resistance bridge based on the cryogenic current comparator
Author :
Seppä, Heikki ; Satrapinski, Alexandre
Author_Institution :
VTT Automation Meas. Technol., Espo, Finland
fDate :
4/1/1997 12:00:00 AM
Abstract :
This paper describes a low-frequency ac resistance bridge based on a Cryogenic Current Comparator (CCC). The bridge is a main tool for resistance traceability of 10 kΩ, 100 Ω, and 1 Ω to the quantum Hall resistance (QHR). The bridge is typically operated with a PC generating a sinusoidal ac signal ranging from (0.1 to 2) Hz. Using a vector analyzer as a signal source the bridge is used to 1 kHz. Both sinusoidal and pseudo random signals have been used to compare the frequency dependence of standard resistors against the QHR. At present the bridge can be used to perform accurate calibrations to 10 Hz, at higher frequencies the self-resonance of the CCC increases the uncertainty
Keywords :
SQUIDs; bridge instruments; calibration; cryogenics; current comparators; electric resistance measurement; measurement errors; quantum Hall effect; resistors; 0.1 Hz to 1 kHz; 1 ohm to 10 kohm; AC resistance bridge; SQUIDs; calibrations; cryogenic current comparator; pseudo random signals; quantum Hall resistance; resistance traceability; self-resonance; standard resistors; uncertainty; vector analyzer; Bridges; Cryogenics; Detectors; Electrical resistance measurement; Frequency dependence; Frequency measurement; Low-frequency noise; Resistors; SQUIDs; Signal generators;
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on