DocumentCode :
1456019
Title :
Line-reflect-match calibration technique for the dual six-port automatic network analyzer
Author :
Ginley, Ronald A.
Author_Institution :
Nat. Inst. of Stand. & Technol., Boulder, CO, USA
Volume :
46
Issue :
2
fYear :
1997
fDate :
4/1/1997 12:00:00 AM
Firstpage :
523
Lastpage :
526
Abstract :
A newly developed method allows dual six-port automatic network analyzers (ANA´s) to be calibrated with a single, known one-port termination instead of air line standards. This technique is especially useful for calibrations below 30 MHz, where air lines cannot be adequately characterized. The choice of the value of the standard is also discussed
Keywords :
calibration; microwave reflectometry; network analysers; 30 MHz; air line standards; calibration; dual six-port automatic network analyzer; line-reflect-match calibration; one-port termination; standard; Bridges; Calibration; Connectors; Equations; Frequency; Measurement standards; Reflection; Skin; Standards development; Testing;
fLanguage :
English
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9456
Type :
jour
DOI :
10.1109/19.571901
Filename :
571901
Link To Document :