DocumentCode
1456019
Title
Line-reflect-match calibration technique for the dual six-port automatic network analyzer
Author
Ginley, Ronald A.
Author_Institution
Nat. Inst. of Stand. & Technol., Boulder, CO, USA
Volume
46
Issue
2
fYear
1997
fDate
4/1/1997 12:00:00 AM
Firstpage
523
Lastpage
526
Abstract
A newly developed method allows dual six-port automatic network analyzers (ANA´s) to be calibrated with a single, known one-port termination instead of air line standards. This technique is especially useful for calibrations below 30 MHz, where air lines cannot be adequately characterized. The choice of the value of the standard is also discussed
Keywords
calibration; microwave reflectometry; network analysers; 30 MHz; air line standards; calibration; dual six-port automatic network analyzer; line-reflect-match calibration; one-port termination; standard; Bridges; Calibration; Connectors; Equations; Frequency; Measurement standards; Reflection; Skin; Standards development; Testing;
fLanguage
English
Journal_Title
Instrumentation and Measurement, IEEE Transactions on
Publisher
ieee
ISSN
0018-9456
Type
jour
DOI
10.1109/19.571901
Filename
571901
Link To Document