• DocumentCode
    1456019
  • Title

    Line-reflect-match calibration technique for the dual six-port automatic network analyzer

  • Author

    Ginley, Ronald A.

  • Author_Institution
    Nat. Inst. of Stand. & Technol., Boulder, CO, USA
  • Volume
    46
  • Issue
    2
  • fYear
    1997
  • fDate
    4/1/1997 12:00:00 AM
  • Firstpage
    523
  • Lastpage
    526
  • Abstract
    A newly developed method allows dual six-port automatic network analyzers (ANA´s) to be calibrated with a single, known one-port termination instead of air line standards. This technique is especially useful for calibrations below 30 MHz, where air lines cannot be adequately characterized. The choice of the value of the standard is also discussed
  • Keywords
    calibration; microwave reflectometry; network analysers; 30 MHz; air line standards; calibration; dual six-port automatic network analyzer; line-reflect-match calibration; one-port termination; standard; Bridges; Calibration; Connectors; Equations; Frequency; Measurement standards; Reflection; Skin; Standards development; Testing;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/19.571901
  • Filename
    571901