DocumentCode :
1456064
Title :
High-frequency (>50 MHz) medical ultrasound linear arrays fabricated from micromachined bulk PZT materials
Author :
Changgeng Liu ; Qifa Zhou ; Djuth, Frank T. ; Shung, K. Kirk
Author_Institution :
Geospace Res. Inc., El Segundo, CA, USA
Volume :
59
Issue :
2
fYear :
2012
fDate :
2/1/2012 12:00:00 AM
Firstpage :
315
Lastpage :
318
Abstract :
This paper describes the development and characterization of a high-frequency (65-MHz) ultrasound transducer linear array. The array was built from bulk PZT which was etched using an optimized chlorine-based plasma dry-etching process. The median etch rate of 8 μm/h yielded a good profile (wall) angle (>;83°) and a reasonable processing time for etch depths up to 40 μm (which corresponds to a 50-MHz transducer). A backing layer with an acoustic impedance of 6 MRayl and a front-end polymer matching layer yielded a transducer bandwidth of 40%. The major parameters of the transducer have been characterized. The two-way insertion loss and crosstalk between adjacent channels at the center frequency are 26.5 and -25 dB, respectively.
Keywords :
biomedical ultrasonics; micromachining; piezoelectric materials; sputter etching; ultrasonic transducer arrays; acoustic impedance; backing layer; crosstalk; etch depths; frequency 50 MHz; frequency 65 MHz; front-end polymer matching layer; high-frequency medical ultrasound linear arrays; median etch rate; micromachined bulk PZT materials; optimized chlorine-based plasma dry-etching process; two-way insertion loss; ultrasound transducer linear array; Acoustics; Etching; Frequency measurement; Materials; Plasmas; Transducers; Ultrasonic imaging;
fLanguage :
English
Journal_Title :
Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
Publisher :
ieee
ISSN :
0885-3010
Type :
jour
DOI :
10.1109/TUFFC.2012.2193
Filename :
6156835
Link To Document :
بازگشت