• DocumentCode
    1456125
  • Title

    Determination of mismatch factor with RF power calibration using air-line impedance standards

  • Author

    Stumper, Ulrich

  • Author_Institution
    Phys. Tech. Bundesanstalt, Braunschweig, Germany
  • Volume
    46
  • Issue
    2
  • fYear
    1997
  • fDate
    4/1/1997 12:00:00 AM
  • Firstpage
    531
  • Lastpage
    534
  • Abstract
    A method to calibrate a coaxial rf power meter by comparison with a standard power meter is described, the ratio of the mismatch factors being determined using two calculable coaxial air-line impedance standards. An initial investigation of the uncertainty of calibration has been made, and initial experimental results of measurements performed with the 7 mm coaxial system and for type N connectors at frequencies from 2 GHz to 18 GHz are presented
  • Keywords
    calibration; electric impedance measurement; impedance matching; measurement standards; microwave measurement; power measurement; radiometry; 2 to 18 GHz; 7 mm; RF power calibration; air-line impedance standards; coaxial impedance standards; coaxial rf power meter; coaxial system; mismatch factor; standard power meter; uncertainty; Calibration; Coaxial components; Impedance; Measurement standards; Performance evaluation; Phase measurement; Power measurement; Radio frequency; Reflection; Wavelength measurement;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/19.571903
  • Filename
    571903